PART A. CALIBRATION CONSTANTS VERIFICATION |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
CALIBRATION VERSION CURRENT? : |
|
YES |
Initials _____ |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
METER MODEL : |
|
|
METER S/N : |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
SCALE MODEL : |
|
|
SCALE S/N : |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
*Current FGIS-approved meters: DJ GAC2500-UGMA, DJ GAC2700-UGMA, and PERTEN AM5200-A. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
PART B. WEIGHING ACCURACY TEST |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
DROP 1 |
DROP 2 |
DROP 3 |
DROP 4 |
DROP 5 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
INSTRUMENT WEIGHT (0.1g) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
SCALE WEIGHT (0.1g) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
INSTRUMENT WT minus SCALE WT |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
AVG OF DIFF: |
|
|
RANGE OF DIFF: |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
WEIGHING ACCURACY TOLERANCE: ±0.5 |
|
|
RANGE TOLERANCE: 1.0 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
PART C. GRAIN MOISTURE SAMPLE TEST |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
SAMPLE TRN: |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
WEIGHT RECORDED ON SAMPLE BAG (0.1 g) |
|
WEIGHT TOLERANCE ± 0.5 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
SCALE WEIGHT OF SAMPLE AND BAG (0.1 g) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
0 |
0 |
0 |
0 |
0 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
INSTRUMENT DATA |
DROP 1 |
DROP 2 |
DROP 3 |
DROP 4 |
DROP 5 |
AVERAGE |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
DISPLAY MOISTURE (0.01%) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
TEST WEIGHT (0.1 lb/bu) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
SAMPLE TEMPERATURE (0.1°C) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
AVG MOISTURE |
|
RANGE OF MOISTURES |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
STANDARD AVG MOISTURE |
|
RANGE OF TEST WEIGHTS |
|
|
OPERATOR (FIELD): |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
DEVIATION |
|
RANGE OF TEMPERATURE |
|
|
DATE TESTED: |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
TOLERANCE |
0.15 |
|
|
|
REVIEWER (HQ): |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
PASS |
FAIL |
|
DATE REVIEWED: |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
SCALE AVERAGE |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
SCALE RANGE |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
MOISTURE AVERAGE |
|
|
MOISTURE RANGE TOLERANCE: 0.26 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
MOISTURE RANGE |
|
|
NO TOLERANCE ON TEST WEIGHT |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
TEMPERATURE |
|
|
TEMPERATURE LIMIITS: 15-27C, 60-80F |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
RECOMMENDED ACTION: |
APPROVED |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
COMMENTS: |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Instructions for Completing Form FGIS-923, “UGMA Moisture Meter Test” |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Indentification Block. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
1. Fill in Field Office or Agency information. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
2. Check the appropriate boxes for the status of check testing the meter. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Part A. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
1. Verify current official calibration version. Update all obsolete calibrations. Initial to indicate that all calibrations are correct. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
For current calibration versions refer FGIS Directive 9180.61. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
2. Enter meter model and serial number along with scale model and serial number. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
3. Enter meter serial number. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
4. Enter SCD1 and SCD2 values. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Part B. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
1. Enter the result of the first drop (meter value). Record the result to the nearest 0.1. |
|
|
|
|
|
|
|
|
|
|
|
|
|
th |
2. Enter the weight reading from the lab scale. Record to result to the nearest 0.1. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
3. Enter the difference (meter minus scale weights), if not calculated automatically. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
4. Repeat Item 1 through Item 4 for a total of five drops. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
5. Enter the average difference (average of five differences of meter minus scale weights), if not calculated automatically. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
6. Enter the range of differences (total spread of the differences of meter minus scale weights), if not calculated automatically. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Part C. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
1. Enter the TRN (sample ID) listed on the sample bag. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
2. Enter the weight recorded on the sample bag to the nearest 0.1. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
3. Remove the wheat sample bag from the outer polyethylene bag and weigh the sample bag on the scale to the nearest 0.1. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
4. Enter the weight reading from the lab scale to the nearest 0.1. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
5. Enter the difference (bag minus scale weight) if not calculated automatically. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
6. Enter the values for information displayed from the first drop on the meter. Moisture, Test Weight, Sample Temperature. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
7. Repeat step 6 for a total of six drops. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
8. Enter the average (average of 5 drops), if not calculated automatically. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
9. Enter the deviation ( average moisture from 5 drops minus Standard avg moisture), if not calculated automatically. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
10. Enter name of the field test meter operator. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
11. Enter the date. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
12. Check the box, if the meter is approved for use, or start the retest or repairing process. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
13. Any additional comments necessary concerning the test or information about the meter. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|